![GE LOGIQ S7 Expert Service Manual Download Page 146](http://html.mh-extra.com/html/ge/logiq-s7-expert/logiq-s7-expert_service-manual_128037146.webp)
GE H
EALTHCARE
DRAFT
LOGIQ™ S7 E
XPERT
/P
RO
D
IRECTION
5460683
, R
EVISION
3
DRAFT (J
UNE
25, 2014)
S
ERVICE
M
ANUAL
Chapter 4 - Functional Checks
4-24
4-3-7-3
Color Flow-Mode
1.) Press CF Mode button.
2.) Increase CF Gain until noise floor shows up.
3.) Verify no vertical “streaks” of color that significantly contrasts the surrounding color.
Figure 4-22 Color Flow-Mode
4-3-7-4
Pulse Doppler Mode
Figure 4-23
1.) From Utility -> Imaging -> General, verify check on “Simultaneous”.
2.) Press B Mode button.
3.) Press PW Mode Button.
4.) Verify that system scan B/PW duplex with the activated probe.