GE MEDICAL SYSTEMS
LOGIQ 200 PRO Series PROPRIETARY MANUAL
2242594
DIAGNOSTICS
2--35
REV 0
2--3--2 System Diagnosis (Continued)
TEST NAME: KEYBOARD TEST
PURPOSE
To check KEYBOARD and interface to MSTE.
CRITERIA
Refer to ILLUSTRATION 2--19 as result of key board test. The latest key is displayed on top line. If
function operate normally, the beep sound.
METHOD
Check a couple of key switch.
Check Trackball.
Check three Rotary Encoders.
Check eight TGC slide ports.
DISPLAY FORMAT OF RESULTS
Service Utility MENU : Keyboard test
99/03/12 09:16:37
Current Software Version is 3.0
Name
Data
FRZ
TB
05,99
FRZ
Gain
-3
BS
TGC
56,33,127,30,00,20,128,54
Set
Set
Clear
Comment
D
c
KEYBOARD TEST MENU
ILLUSTRATION 2--19
Action if the test fails
Change Keyboard assy.
The following Assemblies are necessary for this test
RBF, TBF, SCAN, MVP, MSTE
Summary of Contents for LOGIQ 200