2–3–2 System Diagnosis (Continued)
TEST NAME: SYSTEM DRAM TEST
PURPOSE
To check DRAM in MST assy.
CRITERIA
Refer to result shown on the monitor.
Action if the test fails
Change MST board assy.
The following Assemblies are necessary for this test
RTC, ESP, CONN, DSC, MST, Monitor
MST
RS232C
Controller
con
DRAM
CPU
INTEL 386 EX
Voltage
Supervisor
DIP
SWITCH
Real
Counter
Graphic
Memories
Graphic
Processor
Time
Clock
Program
Memories
SRAM
CNT
Buffer
Quasi–RS232C
9600 BPS
NOVA BUS
PIO BUS
Graphic Controller
To
Rear
Panel
Flash
Memory
for In-site
Peripheral
SYSTEM DRAM TEST
ILLUSTRATION 2–21