GE MEDICAL SYSTEMS
LOGIQ 200 PRO Series PROPRIETARY MANUAL
2242594
DIAGNOSTICS
2--12
REV 0
2--3--2 System Diagnosis (Continued)
TABLE 2--1
SYSTEM DIAGNOSIS TEST LIST
TEST NAME
TEST DETAIL
TEST TYPE
FAILURE CODE
SYSTEM TEST
AUTO
0, 1000~1005
BPHV TEST
Check +THV
AUTO
212
Check --THV
AUTO
213
Memory TEST
AUTO
324
B MODE FUNCTION TEST
ALM TEST
AUTO
323
B MODE FUNCTION TEST
Coordination Test
AUTO
322
Interpolation Test
AUTO
321
Image Capture Test
AUTO
325
Image Archive Test
AUTO
326
Image Recall Test
AUTO
327
Front End Test
AUTO
328
B MODE CINE TEST
AUTO
330, 331, 1003
M MODE FUNCTION TEST
Communication Test
AUTO
341
XY memory Test
AUTO
342
Interpolation Test
AUTO
343
Path Test
AUTO
344
Cine Test
AUTO
345
MVP HARDWARE TEST
Line Memory Test
AUTO
350
Frame Average Memory Test
AUTO
351
B XY Memory Test
AUTO
352
Capture Memory test
AUTO
353
FPGA Test
AUTO
354
USC Generation Test
AUTO
355
DSP SRAM Test
AUTO
356