GE MEDICAL SYSTEMS
LOGIQ 200 PRO Series PROPRIETARY MANUAL
2242594
DIAGNOSTICS
2--29
REV 0
2--3--2 System Diagnosis (Continued)
TEST NAME: B MODE FUNCTION TEST
PURPOSE
To check Acoustic Line Memory, Interpolation, Coordination Transformer and X--Y Memory.
CRITERIA
Refer to result shown on the monitor.
Action if the test fails
If RBF operates properly, change MVP board assy.
The following Assemblies are necessary for this test
Probe, RBF, MVP, MSTE, SCAN, Monitor
MVP
Buffer
FRONT_END
Image
Softener
Frame
Average
PEAK
HOLD
B/M
Front_end
Control
ALM
ITP
XY
memory
CRDTF
ZOOMER
XYADDR
Gen.
Non--Int
Convert
B MODE
DSP BLOCK
CINE
POST
CINE
Control
CINE
Memory
Image
Storage RAM
Image
Display Control
FIFO
USC BUS
I/F
DSP
Program
Memory
XY
Memory
Video
DAC
Video
DAC
Video
Lut
Interlace
Convert
Video
Decoder
Sync
Generator
Buffer
From RBF
SAP
To NOVA_BUS
B MODE FUNCTION TEST
ILLUSTRATION 2--14