2–3–2 System Diagnosis (Continued)
TEST NAME: M MODE FUNCTION TEST
PURPOSE
To check M mode image display control circuit.
CRITERIA
Refer to result shown on the monitor.
Action if the test fails
If RTC operates properly, change DSC board assy.
The following Assemblies are necessary for this test
Probe, RTC, DSC, MST, CONN, Monitor
AAF
ADC
DYN
Image
Softener
Frame
Average
Front_end
M_MODE
XYM
IPT
ALM
CDTRF
B_MODE
Image
/overlay
Selector
RAMDAC
Video
Filter
DSC FRONT_END
Buffer
Buffer
Syn
Generator
DSC
FIFO
ECG
signall
Interface
DSP
320C5I
Program
Memory
XYM
MONITOR
CNT
NOVA
BUS
AR. OL
From
ESP
RTC
BUS
CINE
Controller
Memory
CINE
(Option)
DSC TEST
ILLUSTRATION 2–19