GE MEDICAL SYSTEMS
LOGIQ 200 PRO Series PROPRIETARY MANUAL
2242594
DIAGNOSTICS
2--14
REV 0
2--3--2 System Diagnosis (Continued)
TABLE 2--2
DIAGNOSIS TEST DETAIL
RESULT CODE
Board
Meaning
Passed
Checked
Skipped
The DRAM test, at booting, is skipped.
No PIOP
ECG Module is absent.
212
BPHV
+THV of BPHV is failed.
213
--THV of BPHV is failed.
321
MVP
B Mode interpolation test is failed.
322
B Mode coordination test is failed.
323
B Mode ALM test is failed.
325
Image Capture test is failed.
326
Image Archive test is failed.
327
Image Recall test is failed.
328
Front End test is failed.
330
The first Cine memory test is failed.
331
The Second Cine memory test is failed.
341
M Mode Communication test is failed.
342
M Mode XY memory test is failed.
343
M Mode interpolation test is failed.
344
M Mode path test is failed.
345
M Cine test is failed.
350
Front End Line Memory test is failed.
351
Frame average memory test is failed.
352
B--Mode XY memory test is failed.
353
Image Capture Memory test is failed.
354
FPGA memory test is failed.
355
USC Bus generation test is failed.
356
DSP SRAM test is failed.