DIAGNOSTICS
2–31
REV 2
2–3–2 System Diagnosis (Continued)
TEST NAME: KEYBOARD TEST
PURPOSE
To check KEYBOARD and interface to MST.
CRITERIA
Refer to ILLUSTRATION 2–22 as result of key board test. The latest key is displayed on top line. If
function operate normally, the beep sound.
METHOD
Check a couple of key switch.
Check Trackball.
Check three Rotary Encoders.
Check eight TGC slide ports.
DISPLAY FORMAT OF RESULTS
Service Utility MENU : Keyboard test
96/03/12
09:16:37
Current Software Version is 2.0
Name
Data
FRZ
TB
05,99
FRZ
Gain
–3
BS
Press ’Ctrl + X’ to Exit
KEYBOARD TEST MENU
ILLUSTRATION 2–22
Action if the test fails
Change Keyboard assy.
The following Assemblies are necessary for this test
RTC, ESP, CONN, DSC, MST