GE MEDICAL SYSTEMS
LOGIQ
a
200 PROPRIETARY MANUAL
2138854
DIAGNOSTICS
2–40
REV 2
2–3–6 Initialize
Procedure
Follow these steps to initialize the data which is saved in the LOGIQ
a
200.
Note
If this function is performed, the data which is saved in the LOGIQ
a
200 will be initialized. All preset
parameters including customer presets and system presets will also be set to the factory default.
1.
Using the Arrow up/down keys or Trackball, select the “Initialize” menu in the Top Menu. Refer to the
ILLUSTRATION 2–5.
2.
Press Set or Return key.
The Initialize Menu will appear on the monitor as shown in the ILLUSTRATION 2–29.
Service Utility MENU :
Top Menu
96/03/12
09:16:37
Current Software Version is 2.0
System Diagnosis
System Reset
Self Operation History
Option Set
*Initialize :1 (1:America 2:Europe 3:Asia 4:Japan)
System Clock Adjust
Exit
INITIALIZE MENU
ILLUSTRATION 2–29