GE MEDICAL SYSTEMS
LOGIQ
a
200 PROPRIETARY MANUAL
2138854
DIAGNOSTICS
2–14
REV 2
2–3–2 System Diagnosis (Continued)
RESULT CODE
Meaning
420
Graphic controller is failed.
421
OS ROM is failed.
422
DRAM skipped.
600
PIOP test is failed.
801
5V for analog test is failed.
802
15V test is failed.
803
–15V test is failed.
804
12V test is failed.
805
–5V test is failed.
806
5V test is failed.
900
MAP STOP is failed.
901
Cine frame change is failed.
990
Power test is failed except : 801~806.
Caution for Use System Diagnosis Test
Restricted conditions : Connect the Probe before performing the system diagnosis, and have to remain the probe dur-
ing the system diagnosis.