Countermeasure
Cause
Phenomenon
Make the exposure time longer
The exposure time is too
short
After exposure, no image is displayed
on the monitor
Install the correct light source
The light source is not correct
Install the correct filter
The filter is not correct
Check the sample position
The sample is not in position
Adjust the focus
The focus is not correct
Connect the USB cable correctly
USB cable connection failure
Exposure cannot be performed even if
the Start button has been pressed
Connect the lens cable correctly
Lens cable connection failure
The LAS High Sens. lens is not recog-
nized
Connect the light source correctly
The light source is not con-
nected correctly
The light source is not recognized
Connect the USB cable correctly
USB cable connection failure
The lifting table does not move the tray
to the correct position
Connect the lens cable correctly
Lens cable connection failure
No button is displayed in Focusing
mode
Adjust the focus manually when using the
wide view lens
The wide view lens is installed
6.5 Message “ImageQuant LAS 4000 hardware error occurred"
SensKey:
*
H SensCode:
****
H is displayed.
If the error persists even after the following countermeasures have been taken, fill in the necessary
items and the error code in the Service report fax sheet at the end of the User Manual, and then contact
your GE Healthcare representative.
Countermeasure
Content
SensCode
SensKey
If the problem remains, restart the ImageQuant
LAS 4000 and ImageQuant LAS 4000 Control
Software
UV exposure error
8000
2
Turn off the ImageQuant LAS 4000 power for
a while. After that, restart the ImageQuant LAS
4000 and ImageQuant LAS 4000 Control Soft-
ware
Temperature adjustment er-
ror
9100
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
69
6 Troubleshooting