5 Maintenance
To ensure optimal performance, the ImageQuant LAS 4000 and accessories should be maintained
regularly. This chapter describes the maintenance that should be performed on a regular basis.
NOTICE
Before performing any maintenance on or inside the ImageQuant LAS 4000, be sure
to turn off the power switch and remove the power plug from the electrical outlet.
5.1 ImageQuant LAS 4000
Use a soft dry cloth to clean the exterior of the ImageQuant LAS 4000. For marks that are hard to
remove, wipe with a soft cloth slightly dampened with a neutral detergent, then wipe with a new clean,
dry cloth.
To clean the interior of the instrument, use a non-fluorescent cleanser for biochemical use. If using a
neutral domestic detergent, the detergent residue may fluoresce when exposing images.
NOTICE
Do not use organic solvents. These may damage the surface.
5.2 Sample trays
After using a sample tray, wash it in water using a soft sponge dipped in a neutral detergent. After
cleaning leave the tray to air dry.
WARNING
Wear gloves while cleaning. There may be residues of hazardous chemicals.
NOTICE
Do not use organic solvents. These may damage the surface.
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
61
5 Maintenance