Countermeasure
Cause
Phenomenon
Turn off the power immediately and
contact your GE Healthcare represen-
tative
Interlock failure
The light source is lit even when
the ImageQuant LAS 4000 door
is opened
6.4 Problems with ImageQuant LAS 4000 Control Software
If the error persists even after the following countermeasures have been taken, fill in the necessary
items and the error code in the Service report fax sheet at the end of the User Manual.
Countermeasure
Cause
Phenomenon
Connect the USB cable correctly to the
camera head and PC
USB cable connection failure
ImageQuant LAS 4000 Control Soft-
ware cannot recognize ImageQuant
LAS 4000
Connect the power cable and camera ca-
ble correctly
Power cable or camera cable
connection failure
Set the ambient temperature to 28°C or
less
The ambient temperature is
too high.
The CCD temperature indication does
not change to READY or its value does
not approach the set temperature (af-
ter a sufficient time – about 15 minutes
– has elapsed since the start of cooling)
Set the cooling temperature of the CCD
properly
The CCD temperature is set
too low
Connect the USB cable correctly
USB cable connection failure
Confirm that nothing is blocking the venti-
lation holes
The camera head ventilation
holes are blocked
Connect the USB cable correctly
USB cable connection failure
No focus image is displayed in the Fo-
cusing mode
Adjust the Brightness in the reading soft-
ware
Brightness of reading soft-
ware is too high or low
Install the lens correctly
The lens is not correctly in-
stalled
Check the iris value
The iris is not correct
Adjust the focus
The image is noticeably out
of focus
Turn the power switch of the ImageQuant
LAS 4000 on and reboot it
The ImageQuant LAS 4000
power has been turned off
during exposure
Even when the set exposure time has
elapsed, the reading operation is not
completed.
68
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
6 Troubleshooting
6.3 Problems with ImageQuant LAS 4000