Countermeasure
Cause
Phenomenon
Check that both left and right Epi light
sources are correctly in place and that
they are of the same type
Light source uneven-
ness
Unevenness is seen on the im-
age
Turn the power off and contact your GE
Healthcare representative
A lamp in the UV tran-
silluminator is broken.
Expose the Flat Frame again
The Flat Frame is not
correctly exposed.
Purchase a new UV trans tray.
Degradation of UV
trans tray
There is noise on an image ex-
posed using UV trans tray
6.2 Strange sounds or odors are perceived
Countermeasure
Cause
Phenomenon
Turn off the ImageQuant LAS 4000 power
switch immediately and contact your GE
Healthcare representative
Camera head failure
The camera head generates a strange
noise
Turn off the ImageQuant LAS 4000 power
switch immediately and contact your GE
Healthcare representative
Camera head failure
The camera head generates a strange
odor
Turn off the ImageQuant LAS 4000 power
switch immediately and contact your GE
Healthcare representative
Camera head failure
The camera head emits smoke
Close the door. The subsequent processing
should start.
The door is open
ImageQuant LAS 4000 beeps on and
off (four times, at intervals of three
minutes) and the Busy LED blinks
66
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
6 Troubleshooting
6.1 Problems with the exposed image