5 Holding the latch with one hand, remove the Epi light by pulling it gently outwards.
6 Lift up and remove the Epi light.
7 Before installing the new Epi light, ensure that the latch is sprung open (upwards).
8 Hook the hanger located on top of the Epi light to the top edge of the Epi light base.
1
2
Action
No.
Two guide holes are located on the back of the Epi light base.
1
Slide the hook after attaching the guide pins to the guide holes.
2
46