Program settings button
Countermeasure
Meaning
Error message
Remove some program tem-
plate files that are no longer
required
There are too many exist-
ing program template files
Cannot save.
The maximum number of registered
program templates has been
reached.
Change the filename or choose
to overwrite the older file
A file with the same name
already exists
A program template with the same
name already exists. Do you want to
overwrite it?
Supply a valid file name
Filename starts with an in-
valid character
You cannot specify the first character
of template name as '.' or '-'. Please
start with another character.
The filename is blank
Please input a filename.
Change the wait and exposure
times to appropriate values
The wait time or exposure
time is not appropriate for
the specified image
The wait time or exposure time of
image xx is out of range.
Please change.
Method/Tray Position button
Countermeasure
Meaning
Error message
Ensure that the light source is
installed correctly
The light source could not
be identified
Cannot identlify the light source.
Please check that the light unit is in-
stalled correctly.
Make sure the correct filter for
the method is in place and
registered in the software
The appropriate filter is not
in place
Cannot find the appropriate filter for
the method. Please install an appro-
priate filter.
Option menu - FlatFrame Calibration
Countermeasure
Meaning
Error message
Close the door again and per-
form the flat frame calibration
again
The door was opened dur-
ing flat frame calibration
The door was opened during expo-
sure. Flat frame calibration failed.
Check that the correct sample
tray and calibration plate are
in place and re-calibrate
Flat Frame calibration im-
age exposure failed
An error occurred during flat frame
calibration for Cy3 tray position 4.
Please check that the correct tray
and calibration plate were used.
76
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
6 Troubleshooting
6.6 Error messages and countermeasures
Summary of Contents for ImageQuant LAS 4000
Page 1: ...GE Healthcare Getting Started with ImageQuant LAS 4000 Original instructions...
Page 2: ......
Page 85: ......