Chapter 10 - Care & Maintenance
10-3
How often should care & maintenance tasks be performed?
Table 10-1 Customer Care Schedule
Item
Service at Indicated Time
D
W
M
A
Notes
Air Filter Grid
Remove the filter grid and clean the air filter
grid with vacuum cleaner from outside.
•
more frequently depending on
your environment
Air Filter Grid
Remove filter grid, back top cover and back
cover and clean the housing from inside.
(vacuum cleaner and soft brush)
•
more frequently depending on
your environment
AC Mains Cable
Inspect AC Mains Cable
•
Mobile system check weekly
Cables and Connectors
Check if all cables are fixed well seated at
the correct position and if there is no
mechanical damage visible.
•
also after corrective
maintenance
User Interface
Clean alphanumerical keyboard, Functional
keys, Digital potentiometers, TGC-Shift
potentiometers. (vacuum cleaner, lukewarm
soap water on a soft, damp cloth)
•
Be careful not to get the cloth
too wet so that moisture does
not enter the loudspeakers,
TGC-Slider, or other keys!
LCD Monitor, Touch
Panel and Probe holder
Clean LCD Monitor surface, Touch Panel
and Probe holder with a fluid detergent in
warm water on a soft, damp cloth.
•
Be careful not to get the cloth
too wet so that moisture does
not enter the entire system.
Mechanical parts
Clean and inspect the mechanical function
of wheels, casters, brakes and swivel locks
as well as side door, foot rest, front and rear
handle, and monitor holder.
Remove Dust and Coupling gel.
•
Mobile system check daily
Control Console
movement
Check Translation/Rotation and Height
Adjustment (Elevation)
•
more frequently at Mobile
system
Trackball Check
Check proper operation
(Cursor movement X, Y direction)
•
If failure occurs go to trackball
cleaning.
Trackball Cleaning
Remove trackball ring; open the trackball
housing and take out the trackball
(refer to
Clean the trackball (and X,Y axes) with soft
tissue and screwdriver shaft.
•
Disk Drives
(Data Backup)
Test Image filing (Archive) Import and
Export data capability (DVD/CD Drive)
• •*
* save the image filing data
weekly or at least monthly on
DVD/CD depending on the
number of examinations
Safe Probe Operation
Clean probes and probe cables and check
acoustic lens housing (cracks) and probe
cables. In case of mechanical damage,
don’t use them!
Danger:
Safety risk for
operator and patient.
•*
* or before each use
Probe Air bubbles
To detect air bubbles in filling liquid, shake
the probe carefully and check abnormal
noise.
Probe connectors
Remove dust/dirt of all probe connectors.
Clean with vacuum cleaner if dust is visible.
•
Console Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.