Section 7-5 - Remote Access to Service Platform
Section 7-5
Remote Access to Service Platform
7-5-1
General
This offers GE technicians the possibility to view the entire customer´s desktop and operation system.
Using VCO (Virtual Console Observation) a service technician or the OnLine Center can access and
modify all PC settings and programs or run diagnostics on the customer´s ultrasound scanner.
Remote access to the Voluson® E8 / Voluson® E6 scanner requires permission and customer input
before a GE service technician or OLC can access the customer’s ultrasound scanner remotely.
“Disruptive Mode” can be selected by the customer directly on the Voluson E8 / Voluson E6 ultrasound
system (see:
), or remotely by the service technician or OLC.
7-5-2
How the Customer enables/disables Disruptive Mode and VCO
1.) If not already in read mode,
FREEZE
the image.
2.) Move the cursor to the GE icon and press the
right trackball key
(= right-click).
3.) Select
Connect Clinical Lifeline
(see:
This activates “Disruptive Mode” and “VCO” for the application OLC to quickly assist the customer.
NOTE:
Visual information about the network status is shown in the status area on the bottom of the screen.
4.) If you select
Cancel
, “Disruptive Mode” and “VCO” is turned OFF.
NOTICE
!! NOTICE:
Remote access is
ONLY possible if the service platform is properly configured
(either by the user
or a GE technician at site). Operation see:
Section 7-5-4 "CSD: Configuration" on page 7-15
.
Figure 7-14 Connect Clinical Lifeline
NOTICE
!! NOTICE:
The ACTIVATE SERVICE button in in the Service Tools page (see:
activates/deactivates “Disruptive Mode” and “VCO”.
Gray
=
Idle State
Remote access is inactive.
Green
=
Active State
Disruptive Mode and Virtual Console Observation (VCO) is enabled, but there are no
active incoming connections.
Red
=
Disrupted State
Remote Access is active. All processes [UL_VNC and UL_CSD] are active.
In this state the Voluson E8 / Voluson E6 system should NOT be used clinically.