![GE H48671KK Service Manual Download Page 281](http://html.mh-extra.com/html/ge/h48671kk/h48671kk_service-manual_128609281.webp)
GE H
EALTHCARE
DRAFT
V
OLUSON
E8 / V
OLUSON
E6
D
IRECTION
KTD102576, R
EVISION
7
DRAFT (A
UGUST
23, 2012)
S
ERVICE
M
ANUAL
Chapter 7 - Diagnostics/Troubleshooting
7-9
Section 7-4
Screen Captures and Logs
There may be times when the customer or field engineer will want to capture a presentation on the
screen so it may be recovered by the OnLine Center. This is accomplished by saving the image(s):
A.) to Archive and export them (as jpg, bmp or tiff) to DVD/CD+R/RW or external USB drive.
B.) as jpg and bmp to D:\export by pressing the
ALT
+
P
key on the alphanumeric keyboard.
Note:
Successive
ALT
+
P
keystrokes (max. 20) overwrite existing snapshots at destination HDD!
C.) creates one snapshot (Alt-D.bmp) + “Full Backup” of the System state (fullbackup -> fb1) saved on
D:\export by pressing the
ALT
+
D
key on the alpha-numeric keyboard.
7-4-1
Capturing a screen
The following is the generic process to capture any screen from the scanner.
1.) Navigate to, and display the image/screen/volume to be captured.
2.) Press the
P1
,
P2
,
P3
or
P4
key (depending on system configuration) on the control panel and store
the image onto the clipboard (frame on left side of the screen).
NOTE:
A short summary of
P1
,
P2
,
P3
and
P4
key´s configuration is shown in the status area on the screen.
3.) Select the stored image(s) and export them to DVD/CD+R/RW, an external USB drive (optional) or
mapped Network drive (jpg, bmp, tiff or Volume file).
Figure 7-10 summary of keys configuration