Section 7-9 - Error Messages
ViewerConMngr::vAssembleM_DFE: attempt to get ECG_Consumer_2D- or
ECG_Draw_2D-Block or ECG-CalcHR-Block from ECGViewer Objects failed!!
Action: Restart the system, If after restart not ok then call technical support
restart the system
ViewerConMngr::vAssembleTissueDFE : attempt to get ECG_Consumer_2D- or
ECG_Draw_2D-Block from ECGViewer Objects failed!!
restart the system
Volume_dB Out Of Range
press ok and save this user-setting once again
WMF_KoefBlock: SamplePRF too big
restart the system
Write error! Action: use a new CDR, call technical support
use a new CDR
Write Protected cassette! Remove cassette from VCR and put writeable cassette
into drive of VCR.
Remove cassette from VCR and put writeable cassette
into drive of VCR
Write Protected DVD! Remove DVD and put writeable DVD into drive of recorder.
Put writeable DVD into Drive
Wrong disc type! Please enter a data disc and try again.
enter data disc and try again
Wrong key!
restart the system
Wrong Monitor Type detected!
The installed software supports monitor type “MDM100” only.
Please replace the monitor to avoid image quality issues!
contact your GE sales/service representative
XTD - pucBackScaledBImage: Memory allocation failed
restart the system
XTD - pucDSCBImage: Memory allocation failed
restart the system
XTDTrackballCtrlState::No frame to trackball position found, internal failure!!
restart the system
Error Messages
Actions