Section 8-8 - Replacement of Covers
Section 8-8
Replacement of Covers
Voluson E8 / Voluson E6 cover replacement procedures described in the sub-sections:
8-8-1
Replacement of Footrest Cover
8-8-1-1
Manpower
One person, 1 minute
8-8-1-2
Tools
none
8-8-1-3
Preparations
1.) Power Off/Shutdown the system as described in
8-8-1-4
Footrest Cover - Removal Procedure
1.) For unlocking, turn the 2 quick release devices below the footrest 90°.
2.) Remove the footrest cover.
8-8-1-5
Footrest Cover - Installation Procedure
1.) Place the (new) foot rest cover on the original position.
2.) For locking, turn the 2 quick release devices below the foot rest 90°.
Figure 8-17 turn quick release devices 90° and remove footrest cover
quick release devices
footrest cover