Chapter 8 - Replacement Procedures
8-33
Preparing the Isolation Transformer
4.) Push the fuse holder back into position (correct voltage should be shown in upper line; e.g, 230V,
below) and then close fuse protection.
NOTE:
The adjusted voltage is shown in the viewing window (see:
5.) The output voltage must also be changed by the turn switch, which is located on the right side.
For that purpose use an appropriate screwdriver and make sure the transformer is switched off.
-
for an Input voltage range of
100-130V
set position
A
-
for an Input voltage range of
220-240V
set position
B
6.) Assure that the connected loads can be operated with the chosen voltage.
Figure 8-43 change fuses of isolation transformer
Figure 8-44 output voltage setting
Fuse holder
(e.g. 230V)
voltage
(e.g.
230V
)
is shown in
this window
voltage output
selector