Chapter 8 - Replacement Procedures
8-7
System Software - Installation Procedure (FMI from DVD)
9.) Click the FMI FROM DVD button (see:
) for updating the System Software.
10.)Verify that only the DVD drive is connected to the system, then click OK (
below).
11.)To start update procedure click YES (
The system saves Full Backup in silent mode on R:, then it reboots into LINUX.
A silent “Rollback” image from C:\ is stored on R:\.
After executing all LINUX commands, the system reboots again.
NOTE:
If the currently installed software has to be upgraded to a newer version, and the system DVD being
used is
SW 9.0.3 or higher
, a new software specific “Permanent key is required.
•
Enter the appropriate “Permanent Key”
(calculated in OKOS),
select OK and confirm with
ENTER
.
•
If the entered Permanent key is correct, following window appears, confirm with
ENTER
.
Figure 8-4 Verify that USB devices are disconnected, then click OK
Figure 8-5 Yes - start update procedure
Figure 8-6 insert Permanent Key
Figure 8-7 confirm Permanent Key
Note:
This example shows
fictional numbers.