Chapter 4 - Functional Checks
4-41
Load Full System Configuration (Full Backup)
9.) Click the NEXT button and then select YES to start, or NO to cancel the restore procedure.
After restoring the data, the Voluson E8 / Voluson E6 reboots and the application starts again.
10.)Confirm that the date and time are set correctly
4-5-5
Delete Full System Configuration (Full Backup)
1.) Press the
UTILITIES
key on the control panel.
2.) In the “Utilities” menu touch SYSTEM SETUP to invoke the setup desktop on the screen.
3.) On the left side of the screen select BACKUP and then click the
SYSTEM CONFIGURATION
tab.
4.) Click the DELETE button of the “Full System Configuration” group.
5.) Choose the Destination.
6.) Click on the backup to be deleted (additional information is displayed in the table).
7.) Select the DELETE button.
Figure 4-33 Start Restore Backup now?
Figure 4-34 System Configuration Delete
WARNING
WARNING
!! WARNING:
There is no “ UNDO” function for this action!
WARNING
WARNING
!! WARNING:
When clicking “YES”, the
current data on the system
will be permanently replaced
by the data of the backup and
can not be restored!
choose destination
select backup
click [Delete]