Section 7-9 - Error Messages
Section 7-9
Error Messages
NOTICE
!! NOTICE:
If the problem (error message) still exists after performing the described actions, call technical support.
Error Messages
Actions
3D B_RxLines exceeded! Action: restart the system
restart the system
3D B-O_Frames exceeded! Action: restart the system
restart the system
3D C_RxLines exceeded! Action: restart the system
restart the system
3D C-O_Frames exceeded! Action: restart the system
restart the system
530-Probe connected on
Connect another Probe
530-Probe connected on left probe connector (A).
Connect another Probe
530-Probe connected on middle probe connector (B).
Connect another Probe
530-Probe connected on right probe connector (C).
Connect another Probe
530-Probe connected on upper-left probe connector (CW).
Connect another Probe
A USB-drive has been removed without stopping it. Unplugging or ejecting devices
without first stopping them can often cause crashes and lose valuable data.
press Ok, and stop it before removing it next time
AcquMode - incorrect BBC_Wnd pointer
restart the system
Another drive is already assigned to "xx" that is no DVD/CD rec.
Cannot swap assignment.
Use a DVD/CD writer
Aperture Shape Form for 1.5D Array wasn't found !!! Action: restart the system
restart the system
Application setting ApodTable field count mismatch
restart the system
Application setting dynamic set field count mismatch
restart the system
Application setting field count mismatch
restart the system
Application setting RxFrqResponse field count mismatch
restart the system
AVI Save function fails Action: check the connection and Power cable - restart the
machine and try again
see error message
B_Enhance Out Of Range
press ok and save this user-setting once again
B_Gain Out Of Range
press ok and save this user-setting once again
B_Reject Out Of Range
press ok and save this user-setting once again
B_TxFocus - not calculated and B_SHOT_PART_ON
restart the system.
If after restart not ok then call technical support
backward steps calculated in bGetUnitByStamp out of range
restart the system
BBC_Acquisition Wnd - block creation failed
Action: restart the system
restart the system