Programming Site Data
Setting Response Time -
AVRG
Use this option to specify the number of readings that occur before
the meter will respond to a step change in flow rate. In general, the
smaller the number of readings, the less steady the display will
appear. Complete the following steps to set the response time:
1.
Press
[ESC]
,
[ENTER]
,
[ESC]
. The meter enters the user program.
2.
Press [
X
] until
PROG
appears and press
[ENTER]
.
3.
Press [
X
] until the desired channel appears. Press
[ENTER]
at
CH1
or
CH2
.
4.
Press [
X
] until
SETUP
appears and press
[ENTER]
.
5.
Press [
X
] until
AVRG
appears and press
[ENTER]
.
6.
Press [
W
]
and [
X
] to select the response and press
[ENTER]
. For
best results, select
STATS
(statistics). This increases the response
time under steady flow conditions while still allowing a rapid
response time to changes in flow rate.
Procedure Options
After completing the above steps, the user program returns to the
SET UP
window. Do one of the following:
•
To continue regular programming, refer to Appendix A,
Menu
Maps
, to navigate to the desired menu.
•
To leave the user program, press
[ESC]
four times.
Summary of Contents for AquaTrans AT868
Page 8: ...Chapter 1 ...
Page 24: ...Chapter 2 ...
Page 60: ...Chapter 3 ...
Page 66: ...Chapter 4 ...
Page 72: ...Chapter 5 ...
Page 77: ...Chapter 6 ...
Page 88: ...Chapter 7 ...
Page 98: ...Chapter 8 ...
Page 106: ...Appendix A ...
Page 112: ...Appendix B ...
Page 113: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 LVD Compliance B 2 ...
Page 116: ...Appendix C ...
Page 117: ...Data Records Site Data C 1 ...
Page 122: ...Appendix D ...
Page 123: ...Measuring P and L Dimensions Measuring P and L D 1 ...
Page 126: ...Appendix E ...
Page 127: ...Service Record Introduction E 1 Data Entry E 1 Diagnostic Parameters E 3 ...
Page 131: ...Appendix F ...