46
Solarus 950 Owner’s Manual and User’s Guide
8.
Press the button on the gauge momentarily. This will activate the high vacuum
calibration and adjust the gauge to 7.5 x 10
-5
Torr (1 x 10
-4
mbar).
9.
The calibration is complete. Return the cover to the specimen chamber lid
using the four screws.
Maintenance Report
The maintenance report provides a time/date log of all processes run by the sys-
tem, as well as completion and failure statistics.
ƒigure 25
Example maintenance report
Viewing the Maintenance Report
1.
Touch Options.
2.
Touch Service.
3.
Touch Maintenance Report.
Clearing the Maintenance Report History
1.
Touch Options.
2.
Touch Service.
3.
Touch Reset History.
4.
Touch Yes to confirm.