Solarus 950 Owner’s Manual and User’s Guide
O
Air = 1.012
O
Ar = 1.412
O
Ar(75%)/O
2
(25%) = 1.280
O
N
2
= 1.012
To configure Gas 3 for other gas types:
1.
Touch
Options
. The options screen should be displayed.
2.
Touch
Configuration
. The configuration screen should be displayed.
3.
Be sure that
Gas Line 3 Enable
is set to
Yes
.
4.
Touch
Gas Line 3 Name
, touch
Clear
. Enter the name of the gas,
e.g.,
A r / O 2
. Touch
Apply
. This sets the name.
5.
Touch
Gas Line 3 Multiplier
, touch
Clear
. Enter the multiplier for the gas
you have connected, e.g.,
1 . 2 8
, touch
Apply
. This sets the MFC multiplier,
which compensates for using a gas other than the Ar gas the MFC was cali-
brated for at the factory.
6.
Touch
Done
.
N
OTE
:
These multiplier values can be used with Gas 1 (O
2
), but do not use them
with Gas 2 (H
2
).
Purging the Gas Lines
After the gas lines have been connected and tightened, they should be purged to
eliminate contamination and maintain gas purity
before
opening the valves. This
process is controlled by system software. In most cases, this step will only need to
be performed once, unless the gas lines are removed and reattached or otherwise
opened to atmosphere.