CompuScope 1610C
135
DYNAMIC PARAMETERS
Measured using 1 MHz sine wave input at 10 MS/s with amplitude of 95% of full scale on the ±1V range.
Typical values listed below.
SNR:
70
dB
SFDR:
71
dB
SINAD:
66
dB
THD:
-68
dB
ENOB:
11.5
bits
ACQUISITION MEMORY
Data Storage:
In on-board memory
Memory Sizes:
1M, 8M, 128M, 1G (16 bit samples)
Maximum memory
for each channel:
Up to half on-board memory per channel
TRIGGERING
Number of Trigger Inputs: 2 per card
Source:
CH 1, CH 2, EXT or Software
Type:
Analog
triggering
Sensitivity:
± 20% of full scale
Level Accuracy:
± 10% of full scale
Slope:
Positive or Negative, software selectable
Post Trigger Data:
128 points minimum.
Can be defined with a 64 point resolution
EXTERNAL TRIGGER
Impedance:
1
M
Ω
, 30 pF
Input Type:
Single-ended analog
Amplitude:
Absolute Max ±15 V
Voltage Range:
±1 V and ±5 V
Bandwidth:
10
MHz
Connector:
BNC
TRIGGER OUTPUT
Signal Type:
TTL
Active Edge:
Rising (low-to-high)
Synchronization:
Synchronized to sampling clock
INTERNAL CLOCK
Source:
Clock oscillator
Accuracy:
± 50 ppm (0 to 70 degrees Celsius)
EXTERNAL CLOCK
Maximum Frequency:
20 MHz, maximum
Uses 2x decimation filter
Clock must be present during initialization
Minimum Frequency:
2 kHz
Signal Level:
TTL
Required Duty Cycle:
50% +5%, -0% at 20 MHz
Summary of Contents for CompuScope 14100C
Page 6: ......
Page 24: ...18 Installing CompactPCI PXI Master Slave multi card systems Notes ...
Page 30: ...24 Memory organization on CompuScopes except for CS85GC Notes ...
Page 75: ...Driver installation guide 69 Help About CsTest ...
Page 76: ...70 Driver installation guide Notes ...
Page 90: ...84 CompuScope 85GC Notes ...
Page 174: ...168 Notes Notes ...