CompuScope 14100C
113
Internal Sampling Rate
Single Channel Mode (Channel 1 only):
MS/s:
100, 50, 25, 10, 5, 2, 1
kS/s:
500, 200, 100, 50, 20, 10, 5, 2, 1
Dual Channel Mode (Channels 1 and 2 simultaneously):
MS/s:
50, 25, 10, 5, 2, 1
kS/s:
500, 200, 100, 50, 20, 10, 5, 2, 1
Protection:
1 M
Ω
Impedance:
Diode Clamped
50
Ω
Impedance:
No protection
Connector:
BNC
DYNAMIC PARAMETERS
Measured for a single card configuration using 1 MHz sine wave input at 50 MS/s dual channel mode with
amplitude of 95% of full scale on the ±1 V range. Typical values listed below:
SNR:
63
dB
SFDR:
73
dB
SINAD:
60
dB
THD:
-72
dB
ENOB:
10.2
bits
ACQUISITION MEMORY
Data Storage:
In on-board memory
Memory Sizes:
1M, 8M, 128M, 1G (14 bit samples)
Maximum memory depth:
Single channel:
Up to full on-board memory
Dual channel:
Up to half on-board memory per channel
TRIGGERING
Number of Trigger Inputs: 2 per system
Source:
CH 1, CH 2, EXT or Software
Type:
Analog
triggering
Sensitivity:
± 10% of full scale
Level Accuracy:
± 5% of full scale
Slope:
Positive or Negative, software selectable
Post Trigger Data:
128
(256)
points minimum.
Can be defined with a 64
(128)
point resolution in dual
(single)
channel mode.
EXTERNAL TRIGGER
Impedance:
1
M
Ω
, 35 pF
Amplitude:
Absolute Max ±15 V
Voltage Range:
±1 V and ±5 V
Bandwidth:
30
MHz
Coupling:
AC or DC
Connector:
BNC
Summary of Contents for CompuScope 14100C
Page 6: ......
Page 24: ...18 Installing CompactPCI PXI Master Slave multi card systems Notes ...
Page 30: ...24 Memory organization on CompuScopes except for CS85GC Notes ...
Page 75: ...Driver installation guide 69 Help About CsTest ...
Page 76: ...70 Driver installation guide Notes ...
Page 90: ...84 CompuScope 85GC Notes ...
Page 174: ...168 Notes Notes ...