114
CompuScope 14100C
TRIGGER OUTPUT
Signal Type:
TTL
Active Edge:
Rising (low-to-high)
Synchronization:
Synchronized to on-board system clock
INTERNAL CLOCK
Source:
Clock oscillator
Accuracy:
± 50 ppm (0 to 70 degrees Celsius)
EXTERNAL CLOCK
Maximum Frequency:
100 MHz in single-channel mode
50 MHz in dual-channel mode
Minimum Frequency:
40 MHz in single-channel mode
20 MHz in dual-channel mode
Signal Type:
Sine wave
Signal Level:
MIN 1 V RMS
MAX 2 V RMS
Impedance:
50
Ω
Sampling Edge:
Rising
Coupling: AC
Required Duty Cycle:
50% ± 30% in single-channel mode
50% ± 5% in dual-channel mode
MULTIPLE RECORD
Pre-trigger Data:
None
Record Length:
256
(512)
points minimum.
Can be defined with a 64
(128)
point resolution in dual
(single)
channel mode
MULTI-CARD SYSTEMS
Operating Mode:
Master/Slave or Multiple Independent
Number of Cards
Master/Slave:
2, 4, or 6 cards for 1M memory models
2, 3 or 4 cards for 8M memory models
Available upon request: 128M and 1G memory models
Multiple/Ind.:
Limited by the backplane
Max. # Channels
in Master/Slave Mode:
12 at 50 MS/s (for 1M model)
6 at 100 MS/s (for 1M model)
MASTER/SLAVE SYSTEM TRIGGERING
Number of Trigger Inputs: 2 per system
Trigger Source:
CH 1, CH 2, EXT or Software
(Master Card only)
Sensitivity:
± 10% of full scale
Level Accuracy:
± 5% of full scale
Trigger Slope:
Positive or Negative, software selectable
Summary of Contents for CompuScope 14100C
Page 6: ......
Page 24: ...18 Installing CompactPCI PXI Master Slave multi card systems Notes ...
Page 30: ...24 Memory organization on CompuScopes except for CS85GC Notes ...
Page 75: ...Driver installation guide 69 Help About CsTest ...
Page 76: ...70 Driver installation guide Notes ...
Page 90: ...84 CompuScope 85GC Notes ...
Page 174: ...168 Notes Notes ...