53
BASIC OPERATIONS SETTING PROCEDURES
Remote
Stop
Switches
the
behavior
of the
STOP
signal
It switches the logic for determining how the STOP
signal at the remote control will affect the measurement.
0
~
1
Unit
Change
Unit
Change
Changes the units of the Test Results.
When shipped, it is set to ml/min.
If you want to use
「
Pa
」
, set Digit (point of decimal point)
to
「
0000.
」
. Otherwise, this is invalid.
ml/min
Pa/min
ml/sec
Pa/sec
Pa
Cylinder
Cylinder
Settings
If you use this equipment to control a clamp cylinder,
provide the settings for actuations and the status check
function.
SetValue
X X
・
If NG Hold is set to 0
The cylinder control valve is unclamped after the exhaust
process even if a reject evaluation (NG) is made.
・
If NG Hold is set to 1
The cylinder control valve is not unclamped after the exhaust
process.
・
If Limit Switch is set to 0
The clamp position is not detected and the unclamp process
will not be carried out after the exhaust process.
・
If Limit Switch is set to 1
The clamp position is detected then the unclamp process will
carried out after the exhaust process.
00, 10, 01, 11
5/5
Password Password
Sets the password required to input settings.
Default value is “0000”. *
*This password will be invalid when a new password is
set.
0000
~
9999
NG Hold
Limit Switch
Mode
Logic
0
Suspend measurement when STOP
signal is on.
1
Suspend measurement when STOP
signal is off.
Summary of Contents for FL-601 Series
Page 10: ...9 OUTLINE ...
Page 17: ...16 OUTLINE ADDITIONAL EQUIPMENT MEMO ...
Page 18: ...17 PRINCIPLE ...
Page 24: ...23 POWER ON PROCEDURE ...
Page 50: ...49 BASIC OPERATION ...
Page 59: ...58 TESTING OPERATION ...
Page 68: ...67 MAINTNANCE FUNCTIONS ...
Page 92: ...91 OPTIONAL EQUIPMENT ...
Page 108: ...111 SUPPLEMENT ...
Page 112: ...111 SUPPLEMENT ...
Page 116: ...115 SUPPLEMENT Pneumatic Circuit Diagram FL 601UL L M H1Pneumatic Circuit Diagram ...
Page 119: ...118 SUPPLEMENT Pneumatic Circuit Diagram ChargeBypassUnit CBU 600C Exhaust Valve ...