OPTIONAL EQUIPMENT APU UNIT
(4) Recommended piping circuit for work volumes of at least 2L.
(V Specifications)
MASTER
No Product
Name Remarks
Tank Volume 20 L
①
Air Tank
Tank Volume 38L
②
Precision Regulator
③
Pneumatic Regulator
④
Pressure Sensor
Included with APU
⑤
Charge Bypass Unit
⑥
Filter
①
Air Tank
Vacuum Pump
IN
OUT
Source Pressure
②
IN
OUT
Adjust pressure to
300
~
400kPa
Use as the shortest piping with the largest
possible diameter with the APU.
Front Panel
Pressure Sensor Cable
APU Control Cable
WORK
EXH
VALVE SIGNAL Cable
Rear Panel
When using a V specification exhaust bypass unit, the main body of the FL-601
needs to be altered. Please contact your FUKUDA distributor after purchasing the
tester to help to perform this adjustment.
④
⑤
③
WARNING
For test wiring warnings, please
refer to page 35 in the operations
manual
⑥