POWER ON PROCEDURE
Connection to External Equipment
Status
Judging timing
△:Judging point ←→:Judging range
Setting
Contents
Pressure
1
Pressure
2
Balance
1
Balance
2
Detect
(*1)
Exhaust
Judgment
result shown
on
LCD
display
REMOTE
Terminal
Differential
pressure
sensor
indicates less than
–
FS.
D-NG
-NG
Measurement result
≦
-NG
-NG
-NG
-NG<
Measurement result
<+2NG
OK
OK
+2NG≦
Measurement result
<+NG
+2NG
+2NG
+NG≦
Measurement result
+NG
+NG
+NG>
+2NG
Differential
pressure
sensor
indicates more than +FS.
D+NG
+NG
If the slopes of the first 2 sec and
the second 2 sec differ by more than
10Pa/min during the last 4sec of
fitting detection.
F.NG(*2)
F.NG(*2)
*
1 .Fitting testing time is the fitting detection process.
*
2 . Judges only during the fitting testing time.
*
3 . The pressure sensor used for the P.NG judgment is the same as the TD3 work pressure sensor shown on the
Pneumatic Pressure Circuit Diagram on P.117.
*
4 .2NG is ignored if the 2NG setting is set to be zero.
Summary of Contents for FL-601 Series
Page 10: ...9 OUTLINE ...
Page 17: ...16 OUTLINE ADDITIONAL EQUIPMENT MEMO ...
Page 18: ...17 PRINCIPLE ...
Page 24: ...23 POWER ON PROCEDURE ...
Page 50: ...49 BASIC OPERATION ...
Page 59: ...58 TESTING OPERATION ...
Page 68: ...67 MAINTNANCE FUNCTIONS ...
Page 92: ...91 OPTIONAL EQUIPMENT ...
Page 108: ...111 SUPPLEMENT ...
Page 112: ...111 SUPPLEMENT ...
Page 116: ...115 SUPPLEMENT Pneumatic Circuit Diagram FL 601UL L M H1Pneumatic Circuit Diagram ...
Page 119: ...118 SUPPLEMENT Pneumatic Circuit Diagram ChargeBypassUnit CBU 600C Exhaust Valve ...