BASIC OPERATIONS
RS232C
OPERATIONS
2.
Out Data 2 (Detection Process (DET
、
MDET) data, test result transmission)
A differential pressure value (Pa value) is transmitted every 0.1 sec during the DET phase.
The test result data is transmitted during the EXH process, after the testing is complete.
(The format of the text transmitted for the test result is the same as with Out Data 1)
Format of transmitted text (Detection Process (DET, MDET) data)
Differential
Pressure Value
(
Pa
)
CR
6
1
Item
Byte No.
Data side
+00000
0DH
3.
Out Data 3 (Transmits data and test results from the beginning of the CHG1 phase through the
DET (MDET) phase.
A work pressure value (kPa value) is transmitted every 0.1 sec during the CHG1, CHG2, and BAL 1
processes,
and a differential pressure value (Pa value) is transmitted every 0.1 sec during the BAL2 and DET
(MDET) processes.
(The format of the text transmitted for the test result is the same as with Out Data 1)
Transmitted text format Transmitted text format
(Work Pressure value) (Differential Pressure value
)
Work Pressure
Value
(
kPa
)
CR
Differential Pressure
Value
(
Pa
)
CR
6
1
6
1
Item
Byte No.
Data Ex.
+00.0
-
0DH
Item
Byte No.
Data Ex.
+00000
0DH
○
Test value transmission
The
transmitted test value contents include all the setting items for the currently selected
group.(see page 51~53)
Transmitted text format
Delay(sec)
CR Charge1(sec) CR Charge2(sec) CR Balance1(sec) CR
6
1
6
1
6
1
6
1
Item
Byte No.
Data side
000.0s
0DH
000.0s
0DH
002.0s
0DH
001.0s
0DH
Balance2(sec)
CR Detect(sec) CR Exhaust(sec) CR FDetect(sec) CR
6
1
6
1
6
1
6
1
001.0s
0DH
002.0
0DH
001.0s
0DH
030.0s
0DH
WorkVolume(ml
)
CR
Master
volume(ml)
CR +NG(ml/min) CR +2NG(ml/min) CR
7
1
12
1
6
1
6
1
1.000E2
0DH
1.000E2
-----
0DH
+1.000
0DH
+2.000
0DH
-NG(ml/min)
CR +FNG(ml/min) CR -FNG(ml/min) CR UnitChange CR
6
1
6
1
6
1
6
1
-1.000
0DH
+1.000
0DH
-1.000
0DH
ml
-
kPa
0DH
Lead.P(kPa)
CR Test.P(kPa) CR
Fitting Set
CR
Out Data
CR
6
1
6
1
1
1
1
1
+000.
-
0DH
+000.
-
0DH
0
0DH
0
0DH
Remote Stop
CR UnitChange CR
Cylinder
CR
2
1
6
1
2
1
0
-
0DH
ml
-
kPa
0DH
00
-
0DH
ETX
Space
CR
1
1
1
03H
20H
0DH
The transmission process is shown below.
1.
Press the MAINT key at the initial screen.
2.
Input the password and press the ENT key. Select Data Load on page 2/2 using the key.
3.
Press the ENT key.
4.
Press the START key.
Summary of Contents for FL-601 Series
Page 10: ...9 OUTLINE ...
Page 17: ...16 OUTLINE ADDITIONAL EQUIPMENT MEMO ...
Page 18: ...17 PRINCIPLE ...
Page 24: ...23 POWER ON PROCEDURE ...
Page 50: ...49 BASIC OPERATION ...
Page 59: ...58 TESTING OPERATION ...
Page 68: ...67 MAINTNANCE FUNCTIONS ...
Page 92: ...91 OPTIONAL EQUIPMENT ...
Page 108: ...111 SUPPLEMENT ...
Page 112: ...111 SUPPLEMENT ...
Page 116: ...115 SUPPLEMENT Pneumatic Circuit Diagram FL 601UL L M H1Pneumatic Circuit Diagram ...
Page 119: ...118 SUPPLEMENT Pneumatic Circuit Diagram ChargeBypassUnit CBU 600C Exhaust Valve ...