37
POWER ON PROCEDURE Connection to external equipment
■
Connection to external equipment
To ensure safety during wiring procedures
・
Do not touch the I/C board or start wiring procedures while power is being supplied into the
equipment. Doing so may cause an electric shock.
・
Do not start wiring procedures while connected to the air supply source. Large amounts of
compressed air may be emitted and cause injury. Secure a safe operation environment and then
start the procedure.
WARNING
Do not damage the power cord.
・
Do not damage the power cord or apply heat to it.
・
Placing heavy items or pulling on the power cord may cause fire or electric shock.
CAUTION
Wiring Procedure
・
Select appropriate cables depending on the applied signals and their purpose. Ignoring cable
specifications may cause fire or electric shock.
Ground
・
Carry out grounding construction for the tester’s F.G. terminal.(Better than class 3)
Ignoring this may cause fire or electric shock.
・
In order to prevent malfunctions caused by noise interference, the wiring for the input signal within the fixture and
the output signal power source should be assembled separately. Different power source wiring found in the fixture
should also be assembled separately.
・
Using a shield code for the input signal wire is recommended.
○ REMOTE TERMINAL TABLE
No
Signal Name
Input/Output Pin
No.
Signal Name
Input
/Output
1
STOP
Input
No
Signal Name
Input
/Output
33
OK
Output
2
START
Input
19
+COM
Input
34
+NG
Output
3
FITTING
Input
20
+COM
Input
35
-NG
Output
4
BUBL
Input
21
+COM
Input
36
2NG
Output
5
RETRY
Input
22
+COM
Input
37
BNG
Output
6
GROUP 1
Input
23
+COM
Input
38
PNG
Output
7
GROUP 2
Input
24
+COM
Input
39
RUN
Output
8
GROUP 4
Input
25
NC
-
40
RDY
Output
9
GROUP 8
Input
26
NC
-
41
END
Output
10
CAL
Input
27
-COM
Input
42
ERR
Output
11
Psw
Input
28
-COM
Input
43
NC
Output
12
ENG
Input
29
-COM
Input
44
M/R
Output
13
LM 1
Input
30
-COM
Input
45
NC
-
14
LM 2
Input
31
-COM
Input
46
NC
-
15
NC
-
32
-COM
Input
47
NC
-
16
NC
-
48
NC
-
17
DC+24V
Power Output
49
GND
POWER
18
DC+24V
Power Output
50
GND
POWER
APPLIED CONNECTOR:HONDA MR-50LF
APPLIED CABLE: Larger than 0.2SQ
*
NC in the above list means no connection.
Summary of Contents for FL-601 Series
Page 10: ...9 OUTLINE ...
Page 17: ...16 OUTLINE ADDITIONAL EQUIPMENT MEMO ...
Page 18: ...17 PRINCIPLE ...
Page 24: ...23 POWER ON PROCEDURE ...
Page 50: ...49 BASIC OPERATION ...
Page 59: ...58 TESTING OPERATION ...
Page 68: ...67 MAINTNANCE FUNCTIONS ...
Page 92: ...91 OPTIONAL EQUIPMENT ...
Page 108: ...111 SUPPLEMENT ...
Page 112: ...111 SUPPLEMENT ...
Page 116: ...115 SUPPLEMENT Pneumatic Circuit Diagram FL 601UL L M H1Pneumatic Circuit Diagram ...
Page 119: ...118 SUPPLEMENT Pneumatic Circuit Diagram ChargeBypassUnit CBU 600C Exhaust Valve ...