84
MAINTENANCE FUNTIONS STABILITY CONFIRMATION FUNCTION
5. Press the ENT key.
6. Press the START key. (Begins the stability check function process)
7. The stability check function process will be completed in about one minute.
(If there are no errors within the tester’s air pressure circuit, “No error” will be displayed.)
8. Press the STOP key twice. (To return to the initial screen.)
*If an error is detected in the tester’s air pressure circuit, “ERR” will be displayed.
(Please refer to pages 108~109 for error descriptions.)
Stability check
1
Valve Close
2
Press Start Key
Manual
-
0000Pa
-
0
+
FDetect
Stability check
No error
-0.000ml/m
End
M
anual
+000
kPa
Ready GroupNo 00
Stability check
Err306 +NG
+0.150ml/m
End