101
OPTIONAL EQUIPMENT APU UNIT
■
Optional Equipment APU (Pneumatic Regulator)
○
Summary :
By using an optional APU unit (pneumatic regulator), a stable test pressure (charge) can be obtained in a
shortened time.
Use a dedicated APU unit (type number X005) with the FL-601.
See page 13 for details.
○
Installation directions
:
Figure in ( ) is only for use with 70W
○
APU Control Cable Wiring
Connection
Tester Side
(APU SIGNAL)
APU Side
(IN PUT)
Applied Connector
MR-8LM
RO3-PB8F
Remarks
1
―
D +15V
3
―
B Analog
Input
5(Shield Line)
―
E
(Shield Line)
GND
6
―
F
-
15V
Pin Number
7
―
E GND
Applied Connector
(Included)
MR-8LM (Tester Side)
RO3-PB8F (APU Side)
Applied Cable
At least 0.3 SQ, with a shield. Twisted pair cable
Wiring Precautions
1. Use pin numbers 1 and 6 to pair joint wires.
2. Shield joint wire pin number 5
3. Make sure that the length of the cable is as short as possible.
(Cable length within 5m)
An APU connecting cable (length 1.5 m) is included with the purchase of the APU-
□
-X005
SERIES
.
APU-70W
□-□-X005
APU-90W
□-□-X005
APU-130W
□-□-X005
Height170(164)㎜
Height 220 ㎜
165
145
11
5
10
0
4‐φ7
4‐φ6(5)
110(90)
92(80)
70
(
66
)
60
(5
0)
1 6
4
2 7
5
3 8
A G
B H
F
C
D
E
MR-8LM
Rear panel terminal
R03-PB8F
Rear panel terminal
Shield Line
Summary of Contents for FL-601 Series
Page 10: ...9 OUTLINE ...
Page 17: ...16 OUTLINE ADDITIONAL EQUIPMENT MEMO ...
Page 18: ...17 PRINCIPLE ...
Page 24: ...23 POWER ON PROCEDURE ...
Page 50: ...49 BASIC OPERATION ...
Page 59: ...58 TESTING OPERATION ...
Page 68: ...67 MAINTNANCE FUNCTIONS ...
Page 92: ...91 OPTIONAL EQUIPMENT ...
Page 108: ...111 SUPPLEMENT ...
Page 112: ...111 SUPPLEMENT ...
Page 116: ...115 SUPPLEMENT Pneumatic Circuit Diagram FL 601UL L M H1Pneumatic Circuit Diagram ...
Page 119: ...118 SUPPLEMENT Pneumatic Circuit Diagram ChargeBypassUnit CBU 600C Exhaust Valve ...