60
FITTING
TESTING SETTING PROCEDURES
■
Fitting Testing Procedure
1. Press the FITTING key. (It begins reading the Fitting Data)
2. The test result will be displayed when the testing has concluded.
If fitting set 0, 1 or 2 is selected, there will be an indication in the upper-left side of the screen that
tells you if the value was determined freely, or using the slope. If it indicates that the slope was
used, this means that the result of formula 1 on page 20 does not coincide with actual data. If this
occurs, the constants a, b and c from page 20 were not calculated, meaning that even if you press
START to initiate the next measurement, a fitting measurement will be performed first.
F OK
No update
Slope
1.000
ml/m
Exhaust
GroupNo 00
・
Is the AIP SUPPLY(Test Pressure) being supplied?
・
Is the A.P.SUP (Pilot Valve Operation Pressure) being supplied?
・
Is the work bowl valve opened?
・
Is the work connected to the work port?
・
Are each of the setting values correct?
Preparations
Testing Value
Judgment
Result
Indicates that
the unit is
currently
performing
Fitting testing
F Manual
-
0000Pa
-
0
+
FDetect
GroupNo 00
Summary of Contents for FL-601 Series
Page 10: ...9 OUTLINE ...
Page 17: ...16 OUTLINE ADDITIONAL EQUIPMENT MEMO ...
Page 18: ...17 PRINCIPLE ...
Page 24: ...23 POWER ON PROCEDURE ...
Page 50: ...49 BASIC OPERATION ...
Page 59: ...58 TESTING OPERATION ...
Page 68: ...67 MAINTNANCE FUNCTIONS ...
Page 92: ...91 OPTIONAL EQUIPMENT ...
Page 108: ...111 SUPPLEMENT ...
Page 112: ...111 SUPPLEMENT ...
Page 116: ...115 SUPPLEMENT Pneumatic Circuit Diagram FL 601UL L M H1Pneumatic Circuit Diagram ...
Page 119: ...118 SUPPLEMENT Pneumatic Circuit Diagram ChargeBypassUnit CBU 600C Exhaust Valve ...