50
TURBOTEST 500B User’s Guide
Do the following to the TURBOTEST 500B
3 Return to the [OPM] test mode.
4 Attach the APC connector on the reference cable to the OPM Port.
5 Attach an adapter to the connector at the other end of the reference cable.
6 Attach the reference cable on the Optical Light Source to the adapter on the TURBOTEST 500B
reference cable.
7 Press the [
λ]
soft key to select the required wavelength.
8 Press the [MENU] key to access the [OPM MAIN MENU] screen.
9 Use either [UP] or [DOWN] arrow key to highlight the [REFERENCES] option.
10 Press the [SELECT] soft key to access the [OPM REFERENCE] screen.
11 Press the [SET REF] soft key to display the {OPM SET REFERENCE] sub-screen. The reference value
will be displayed on the sub-screen.
12 Press the [SET] soft key to save the displayed reference value.
Once completed, the unit is now measuring optical loss from the Optical Light Source to the
TURBOTEST 500B.
Figure 7-2. TURBOTEST 500B Reference Setup.
TURBOTEST
500B
Optical
Light
Source
Adapter
APC Connector
Reference Cable
4
2
Reference Cable
UPC Connector
OPM Port
5
6
Optical Port
UPC Connector