25
Technical data
Extract: see the Technical Handbook for details
1)
.
Ambient conditions:
during operation and
storage
–20 to +50
°C for NiMH power packs of type: HBT
0000 and HBT 0100,
for Alkaline single cells of type:
Duracell Procell MN 1500
2)
–20 to +40
°C for NiMH single cells of type:
GP 180AAHC
2)
and for Alkaline single cells of type:
Panasonic LR6 Powerline
0 to +40
°C for Alkaline single cells of type:
Varta 4006
2)
, Varta 4106
2)
,
700 to 1300
hPa
10 to 90
% (to 95
% briefly) r.h.
Position of use
any
Storage time
X-am 5600
1 year
Sensors
1 year
Protection class
IP
67 for instrument with sensors
Alarm volume
Typically 90
dB (A) in 30
cm
distance
Operation time
– Alkaline battery/NiMH
rechargeable batteries
(battery holder ABT
0100 (X-am 5600))
Typically 9
hours under standard conditions
- NiMH power pack
T4
(HBT 0000)
T4 HC (HBT 0100)
Typically 9
hours under standard conditions
Typically 10.5
hours under standard conditions
Dimensions
approx. 130 x 48 x 44
mm (H x W x D)
Weight
approx. 220 to 250
g
Refresh interval for
display and signals
1 s
CE marking:
Electromagnetic compatibility
(Directive 2004/108/EC)
Explosion protection (Directive 94/9/EC)
Approvals
3)
:
(see “Notes on Approval” on page
The BVS 10 ATEX E 080 X technical suitability test
is based on the adjustment with the target gas.
1)
The Technical Handbook and Instructions for Use/data sheets for the sensors used and the PC
software CC-Vision for Dräger X-am 5600 are available on the product page of the
X-am
5600 at the following Internet address: www.draeger.com.
See also attached Instructions for Use and Data Sheets of the sensors used.
2)
Not subject to BVS10 ATEX E 080X and PFG 10 G 001X performance approval.
3)
The technical approvals are valid for the X-am 5600 gas detection instrument and the calibration
cradle. The explosion-protection approvals are only valid for the X-am 5600 gas detection
instrument; the calibration cradle must not be used in the Ex zone.
Summary of Contents for Drager X-am 5600
Page 163: ...162 O2 1 Bump Test 21 O2 A1 A1 A1 1 2 OK A2 Dr ger X am 5600 A1 O2 A2 O2 A1 A2...
Page 228: ...227 O2 Bump Test 21 Vol O2 A1 A1 A1 A1 A2 OK A2 OK A1 O2 A2 O2 A1 A2...
Page 292: ...291 9LVLRQ DP 7 67 7 67 67 7 7 7 7 7 U JHU DP 2 m EW m 7 67 6 2 2 2...
Page 293: ...292 9RO 2 U JHU DP 5 2 2...
Page 294: ...293 67 7 67 7 67 2 U JHU 67 7 2 2 7 67 9LVLRQ 2 U JHU 8 5 2 U JHU 5 2 U JHU 6 2 U JHU 9LVLRQ 1...
Page 297: ...296 2 5 5 YRO 5 2 5 2 YRO 2 YRO 2 SSP 6 SSP YRO U JHU 2 2 9RO 6 SSP 2 SSP 2 2 PLQ 2 2 2 2...
Page 298: ...297 1 1 2 6 2 5 2 U JHU U JHU U JHU U JHU U JHU...
Page 302: ...301...
Page 303: ...302...