24
Cleaning
The instrument does not need any special care.
Dirt and deposits can be removed from the instrument by washing it with
cold water. A sponge can be used for wiping if necessary.
Carefully dab dry the instrument using a cloth.
Maintenance
The instrument should be inspected and serviced once a year by suitably
qualified persons. Comparisons:
EN 60079-29-2 – Gas detectors - Selection, installation, use and
maintenance of detectors for flammable gases and oxygen
EN
45544-4 – Electrical apparatus used for the direct detection and direct
concentration measurement of toxic gases and vapours - Part 4: Guide for
selection, installation, use and maintenance
National regulations
Recommended calibration interval for measuring channels O2, H
2
S, H
2
and CO:
6 months.
Recommended calibration intervals for the measuring channel IR Ex/CO
2
: 12
months.
Calibration intervals of other gases: see Instructions for Use of the applicable
DrägerSensors.
See the Technical Manual for details of spare parts.
Storage
Dräger recommends storing the instrument in the charger module (order
no. 83
18
639).
Dräger recommends checking the charge of the power supply at least every
three
weeks if the instrument is not stored in the charger module.
CAUTION
Abrasive cleaning implements (brushes etc.), cleaning agents and cleaning
solvents can destroy the dust and water filters.
Disposing of the instrument
Dispose of the product in accordance with the applicable rules and regulations.
WEEE
Battery disposal
In accordance with EU Directive 2002/96/EC, this product must not be
disposed of as household waste. This is indicated by with the adjacent
icon.
You can return this product to Dräger free of charge. For information
please contact the national marketing organisations and Dräger.
In accordance with EU Directive 2006/66/EC, batteries and
rechargeable batteries must not be disposed of as household waste
but must be taken to battery collection centres. This is indicated by the
adjacent icon.
Collect batteries and rechargeable batteries as specified by the
applicable regulations and dispose of at battery collection centres.
Summary of Contents for Drager X-am 5600
Page 163: ...162 O2 1 Bump Test 21 O2 A1 A1 A1 1 2 OK A2 Dr ger X am 5600 A1 O2 A2 O2 A1 A2...
Page 228: ...227 O2 Bump Test 21 Vol O2 A1 A1 A1 A1 A2 OK A2 OK A1 O2 A2 O2 A1 A2...
Page 292: ...291 9LVLRQ DP 7 67 7 67 67 7 7 7 7 7 U JHU DP 2 m EW m 7 67 6 2 2 2...
Page 293: ...292 9RO 2 U JHU DP 5 2 2...
Page 294: ...293 67 7 67 7 67 2 U JHU 67 7 2 2 7 67 9LVLRQ 2 U JHU 8 5 2 U JHU 5 2 U JHU 6 2 U JHU 9LVLRQ 1...
Page 297: ...296 2 5 5 YRO 5 2 5 2 YRO 2 YRO 2 SSP 6 SSP YRO U JHU 2 2 9RO 6 SSP 2 SSP 2 2 PLQ 2 2 2 2...
Page 298: ...297 1 1 2 6 2 5 2 U JHU U JHU U JHU U JHU U JHU...
Page 302: ...301...
Page 303: ...302...