23
Adjusting the sensitivity for an individual
measuring channel
–
The span adjustment can be carried out specifically for individual sensors.
–
During the span adjustment, the sensitivity of the selected sensor is set to
the value of the used test gas.
–
Use a standard test gas.
–
Allowed test gas concentration:
Connect the test gas cylinder with the calibration cradle.
Vent the test gas into a fume cupboard or into the open air (with a hose
connected to the second connector of the calibration cradle).
Switch on the instrument and insert it into the calibration cradle.
Press the
[
+
]
key and keep it pressed for 5
seconds to call the calibration
menu, enter the password (password on delivery = 001).
Use the
[
+
]
key to select the single gas adjustment function. The symbol for
span adjustment
»
«
flashes.
Press the
[
OK
]
key to start the channel selection.
NOTICE
A potentially activated H
2
offsetting is automatically temporarily deactivated
during a manual adjustment, a PC calibration or an automatic Bump Test for
the relevant duration.
DUAL
IR
Ex
IR
Ex
20 to 100 %LEL
1)
2)
/ 5 to 100
vol.-%
1) 2)
1)
Depending on the selected data set.
2)
Depending on the measuring range and the measuring accuracy.
DUAL
IR
CO
2
IR
CO
2
0.05 to 5
vol.-%
2)
O
2
10 to 25
vol.-%
CO
20 to 999
ppm
H
2
S
5 to 99
ppm
H
2
HC
0.5 to 4.0
vol.-%
Test gas concentration of other gases: see Instructions for Use for the
respective DrägerSensors.
WARNING
Risk to health! Do not inhale the test gas.
Observe the hazard warnings of the relevant safety data sheets.
–
The display flashes the gas of the first measuring channel, e.g.
»
ch
4
- LEL
«
.
Press the
[
OK
]
key to start the adjustment function of this measuring
channel, or use the
[
+
]
- key to select another measuring channel (O
2
- vol.-
%, H
2
S - ppm or CO - ppm, etc.).
–
The test gas concentration is displayed.
Press the
[
OK
]
key to confirm the test gas concentration or use the
[
+
]
key
to change the test gas concentration and complete the process by pressing
the
[
OK
]
key.
–
The measurement value flashes.
Open the test gas cylinder valve to let gas flow over the sensor with a
volume flow of 0.5
l/min.
–
The displayed, flashing measurement value changes to the value according
to the supplied test gas.
When the measurement value displayed is stable (after at least
120
seconds):
Press the
[
OK
]
key to perform the adjustment.
–
The display containing the current gas concentration changes with the
display
»
OK
«
.
Press the
[
OK
]
key or wait about 5
seconds in order to complete the
adjustment of this measuring channel.
–
The next measuring channel may appear for adjustment.
–
After the adjustment of the last measuring channel, the instrument changes
to the measuring mode.
Close the test gas cylinder valve and remove the instrument from the
calibration cradle.
If a fault has occurred during the span adjustment:
–
The fault message »
« appears and »
« is displayed for the
respective sensor instead of the measured value.
In this case, repeat the adjustment.
If necessary, replace the sensor.
NOTICE
In the case of the CO
2
channel there is a 2-stage calibration routine:
The zero point adjustment takes place first and is then followed by the span
adjustment.
Summary of Contents for Drager X-am 5600
Page 163: ...162 O2 1 Bump Test 21 O2 A1 A1 A1 1 2 OK A2 Dr ger X am 5600 A1 O2 A2 O2 A1 A2...
Page 228: ...227 O2 Bump Test 21 Vol O2 A1 A1 A1 A1 A2 OK A2 OK A1 O2 A2 O2 A1 A2...
Page 292: ...291 9LVLRQ DP 7 67 7 67 67 7 7 7 7 7 U JHU DP 2 m EW m 7 67 6 2 2 2...
Page 293: ...292 9RO 2 U JHU DP 5 2 2...
Page 294: ...293 67 7 67 7 67 2 U JHU 67 7 2 2 7 67 9LVLRQ 2 U JHU 8 5 2 U JHU 5 2 U JHU 6 2 U JHU 9LVLRQ 1...
Page 297: ...296 2 5 5 YRO 5 2 5 2 YRO 2 YRO 2 SSP 6 SSP YRO U JHU 2 2 9RO 6 SSP 2 SSP 2 2 PLQ 2 2 2 2...
Page 298: ...297 1 1 2 6 2 5 2 U JHU U JHU U JHU U JHU U JHU...
Page 302: ...301...
Page 303: ...302...