20
After leaving the area, if the concentration is less than the alarm threshold A2:
Press the
[
OK
]
key, the alarm messages are switched off.
STEL / TWA exposure alarm
–
The STEL and TWA alarm cannot be acknowledged or cancelled.
Switch off the instrument. The values for the exposure evaluation are
deleted after the instrument is switched on again.
Battery pre-alarm
Acknowledging the pre-alarm:
Press the
[
OK
]
key, only the audible alarm and the vibration alarm are
switched off.
–
The battery still lasts approx.
20
minutes after the first battery pre-alarm.
Battery main alarm
The battery main alarm cannot be acknowledged or cancelled:
–
The instrument is automatically switched off after 10
seconds.
WARNING
Danger to life! Leave the area immediately.
A main alarm is self-latching and cannot be acknowledged or cancelled.
The alarm is indicated by an intermittent alarm
message:
Display
»
A2
«
and
»
«
(STEL) or
»
«
(TWA) and measured value
alternating:
WARNING
Risk to health! Leave the area immediately.
After this alarm, the deployment of personnel is subject to the relevant national
regulations.
The alarm is indicated by an intermittent alarm
message:
Flashing special symbol
»
«
on the
right side of the display:
The alarm is indicated by an intermittent alarm
message:
Flashing special symbol
»
«
on the right side of the display:
–
Before the instrument is switched off, the visual, audible and vibration
alarms are activated for a short time.
Instrument alarm
–
The instrument is not ready for operation.
Commission maintenance personnel or the Dräger Service to eliminate the
error.
Calling the Info Mode
In measuring mode, press the
[
OK
]
key for approx. 3
seconds.
–
If any warning or fault messages exist, the corresponding note or error
codes are displayed (see the Technical Handbook). Press the
[
OK
]
key
successively for the next display. The peak values and the exposition
values TWA and STEV will be displayed.
–
If no key is pressed for 10
seconds,
the
instrument
returns
automatically
to
measuring mode.
Calling the Info Off Mode
With the instrument switched off, press the
[
+
]
key for approx. 2
seconds.
The name of the gas, measuring unit and measuring range limit value are
displayed for all channels.
Pressing the
[
+
]
key again exits the Info Off Mode (or via timeout).
Calling the Quick Menu
In measuring mode, press the
[
+
]
key three times.
–
If functions in the quick menu are activated using the PC software Dräger
CC-Vision, you can select these functions using the
[
+
]
key. If no functions
have been activated in the quick menu, the instrument remains in measuring
mode.
Possible functions:
1. Bump test
2. Fresh-gas adjustment
1)
3. Delete peak values
The alarm is indicated by an intermittent alarm
message:
Special symbol
»
«
displayed on the right side of the display:
1)
The fresh-gas adjustment/zero-point adjustment is not supported by the DrägerSensor DUAL IR
CO
2
, DrägerSensor IR CO
2
or the DrägerSensor XXS O
3
. A zero-point adjustment of these
sensors can be made with the Dräger CC Vision PC software. A suitable zero gas free from
carbon dioxide and ozone (e.
g.
N
2
) must be used.
Summary of Contents for Drager X-am 5600
Page 163: ...162 O2 1 Bump Test 21 O2 A1 A1 A1 1 2 OK A2 Dr ger X am 5600 A1 O2 A2 O2 A1 A2...
Page 228: ...227 O2 Bump Test 21 Vol O2 A1 A1 A1 A1 A2 OK A2 OK A1 O2 A2 O2 A1 A2...
Page 292: ...291 9LVLRQ DP 7 67 7 67 67 7 7 7 7 7 U JHU DP 2 m EW m 7 67 6 2 2 2...
Page 293: ...292 9RO 2 U JHU DP 5 2 2...
Page 294: ...293 67 7 67 7 67 2 U JHU 67 7 2 2 7 67 9LVLRQ 2 U JHU 8 5 2 U JHU 5 2 U JHU 6 2 U JHU 9LVLRQ 1...
Page 297: ...296 2 5 5 YRO 5 2 5 2 YRO 2 YRO 2 SSP 6 SSP YRO U JHU 2 2 9RO 6 SSP 2 SSP 2 2 PLQ 2 2 2 2...
Page 298: ...297 1 1 2 6 2 5 2 U JHU U JHU U JHU U JHU U JHU...
Page 302: ...301...
Page 303: ...302...