Chapter 4: Operation
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
73
4.10.2.2 Daily report AquaA2
In the daily report, the current data of the device in the
SUPPLY
mode
are recorded every day at a programmed time. These data are intended
to assist the technician when analyzing the device.
Data of the previous day:
– Date and time
– Concentrate flow
FL-C
– Rejection rate
Rej.
– Standardized dialysis water flow
FL-Pstd
– Water consumption of the previous day
– Total water consumption
The following data are recorded:
– Date and time when the data was
collected
– Dialysis water conductivity
CD-Ps
– Dialysis water temperature
T-Ps
– Feed pressure sensor
P-Fs
– Concentrate pressure
P-Cs
– Permeate pressure
P-Ps
– Feed flow
FL-Fs
– Concentrate flow
FL-Cs
– Rejection rate
Rej.