Cooling ring main
Module heat disinfection
Example temperature graph during module heat disinfection:
The temperature graph shows the temperature over time during module
heat disinfection:
If an
AquaA2
is connected, the membranes of the
AquaA2
will be
included in the heat disinfection.
During the first part of this phase, the ring
main is cooled to a temperature below 35 °C
by discarding hot product water.
During the second part of this phase, rinsing
of the ring main is started. Here, dialysis
water is discarded on the
RingBase
for a
preprogrammed duration. This part of the
“Cooling ring main” phase can be stopped
prematurely at any time using the
Cancel
button.