Chapter 12: Specifications
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
181
12.13 Specifications–AquaUF
Only the conditions which deviate from the specifications for the
AquaA
are listed.
Operating conditions
Development of heat
None
Feed water
Dialysis water of the
AquaA
Water inlet temperature
+5 °C to 35 °C (for heat disinfection, the values for the
AquaHT
apply)
Inlet volume AquaA
Rated output
Storage conditions
Storage temperature range
+5°C to +40°C
Relative air humidity
Max. 70 % at 20 °C; non-condensing
Single ultrafilter
Dual ultrafilter
Environmental conditions
Flow
2500 L/h
4000 L/h
(at 15 °C and
Δ
p 1 bar)
Pressure drop
0.7 bar
1.2 bar at 4000 L/h
at 15 °C
Max. inlet pressure
6 bar
6 bar
at 50 °C
Max. inlet pressure
4 bar
4 bar
at 80 °C
Dimensions W/H/D
1600/400/400
1600/400/400
Weight empty/filled
28/35 kg
32/45 kg
Note
The
AquaUF
must be stored in a well-ventilated room with little variation
in temperature.
To prevent bacterial growth, the
AquaUF
must be completely drained
in case of prolonged storage times and particularly in case of increased
storage temperatures.
Note
Protect the device from frost!
Summary of Contents for AquaA
Page 2: ......
Page 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Page 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...