Chapter 5: Alarms
118
Fresenius Medical Care
AquaA
IFU-EN 06A-2021
W–02–01–17*
Signal:
yellow
WARNING:
concentrate
pressure below
alarm limit
– The concentrate pressure is
below the specified alarm
limit.
– Pressure sensor P-C
defective
– Booster pumps do not
deliver or do not build up
any pressure.
– Measuring transducer B4
defective
Contact the service
department
W–02–01–18*
Signal:
yellow
WARNING
Permeate temp.
alarm limit
exceeded
– The dialysis water
temperature T-P has
exceeded the specified
alarm limit during
AquaA2
operation.
– Temperature sensor T-P
defective
– Measuring transducer B4
defective
Contact the service
department
Error code
Visual
indica-
tor
Messages
Cause
Action required