Chapter 9: Installation
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
147
9.2 Operational Qualification
9.2.1
To be observed before Operational Qualification
Tester's qualification
Operational Qualification must be performed by the technical service
department of Fresenius Medical Care or a person authorized by them.
The Operational Qualification may only be performed by persons
qualified to properly perform the specified checks based on their
educational background, training, knowledge, and experience.
Furthermore, the persons performing the checks must not be bound by
any directives when performing this activity.
Only for Operational
Qualification
The following information is only intended for the operational
qualification. It is not applicable for operational requalification of devices
that have been removed from service or temporarily shut down.
Specifications
– Observe the information on the specifications.
– Specific connection and performance data must be taken from the
Specifications chapter.
Electromagnetic radiation
Do not use devices emitting electromagnetic radiation (e.g., walkie-
talkies, mobile phones, radio transmitters) in the vicinity of the device in
operation. This may cause a malfunction of the device.
Power plug
The power plug must be easily accessible.
Use of spare parts
Any installation, modification or repair work requiring the device to be
opened may only be performed by manufacturer-authorized persons
and is permitted only when using original spare parts.
Test equipment and
accessories
The activities described in this document require the availability of the
necessary technical test equipment and accessories.
Precautions
Before turning power on, repair any visible damage.
Prior to opening the device and when working on the open device, the
following precautions must be observed:
Protect the components against ingress of fluids.
Do not touch live parts.
Disconnect and connect all jacks, connectors and components only
when the device is turned off.
ESD precautions
When repairing the system and when replacing spare parts, observe
the applicable ESD precautions.
Summary of Contents for AquaA
Page 2: ......
Page 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Page 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...