Chapter 15: Appendix
236
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
Fig. 5 - rinsing the sampling valve
Turn the multiway valve 90° clockwise.
Rinse the sampling valve for approx. 60 seconds via the rinse
tubing. (
Fig. 5
)
Fig. 6 - filling the bag
Then turn the multiway valve again 90° clockwise to fill the bag.
(
Fig. 6
)
Caution:
Make sure to return the multiway valve in time to its original
position (
Fig. 4
) to prevent bursting of the bag!
Completing the sampling procedure
The sampling valve is now closed again by turning it clockwise.
Clamp of the disposable parts after the multiway valve and
immediately
close the bag with the enclosed plug.
Gently squeeze the bag to check for any leaks.
Affix the completed label on the bag and place it into the
prepared transport box.
Note
Local policy and procedures may supersede the manufacturers
sampling methods.
Illustration
Description
Fig. 5
Fig. 6
Summary of Contents for AquaA
Page 2: ......
Page 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Page 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...